A nitrogen triple-point thermal storage unit for cooling a SQUID magnetometer

Share/Save/Bookmark

Rijpma, A.P. and Meenderink, D.J. and Reincke, H.A. and Venhorst, G.C.F. and Holland, H.J. and Brake ter, H.J.M. (2005) A nitrogen triple-point thermal storage unit for cooling a SQUID magnetometer. Cryogenics, 45 (3). pp. 231-239. ISSN 0011-2275

[img]PDF
Restricted to UT campus only
: Request a copy
1161Kb
Abstract:In order to achieve turnkey operation, we plan to use cryocoolers to cool a SQUID magnetometer system. To minimize the magnetical and mechanical interference from the coolers, we intend to switch them off during the actual measurements. Consequently, a thermal storage unit (TSU) is required with sufficient capacity at an appropriate temperature (<77 K). In a feasibility study, we consider a load of 0.5 W from the SQUID sensor unit and an operating time of 10 h. To account for an increased load caused by the TSU itself, an overall capacity of 15 Wh is aimed at. The nitrogen triple-point is chosen because of the large latent heat involved in the transition from solid to liquid and the corresponding well-suited temperature (63 K). Furthermore, any safety risks involved with the use of nitrogen are small compared to alternatives. To contain the nitrogen, highly porous alumina is used. A structure was made in which layers of copper and porous material alternate, thus establishing a good thermal contact between the nitrogen and the casing of the TSU. Experiments show an overall capacity of the system around 85% of the expected theoretical value. Suggestions for improvements are given so as to arrive at a TSU capacity of 15 Wh.
Item Type:Article
Copyright:© 2005 Elsevier
Faculty:
Science and Technology (TNW)
Research Group:
Link to this item:http://purl.utwente.nl/publications/77647
Official URL:http://dx.doi.org/10.1016/j.cryogenics.2004.10.005
Export this item as:BibTeX
EndNote
HTML Citation
Reference Manager

 

Repository Staff Only: item control page

Metis ID: 224714