Low-noise rotating sample holder for ultrafast transient spectroscopy at cryogenic temperatures

Share/Save/Bookmark

Fanciulli, R. and Cerjak, I. and Herek, J.L. (2007) Low-noise rotating sample holder for ultrafast transient spectroscopy at cryogenic temperatures. Review of Scientific Instruments, 78 (5). 053102. ISSN 0034-6748

open access
[img] PDF
412kB
Abstract:We present the design and testing of a rotating device that fits within a commercial helium cryostat and is capable of providing at 4 K a fresh sample surface for subsequent shots of a 1–10 kHz amplified pulsed laser. We benchmark this rotator in a transient-absorption experiment on molecular switches. After showing that the device introduces only a small amount of additional noise, we demonstrate how the effect of signal degradation due to high fluence is completely resolved