Real-time in vivo photoacoustic and ultrasound imaging

Share/Save/Bookmark

Kolkman, Roy G.M. and Brands, Peter J. and Steenbergen, Wiendelt and Leeuwen van, Ton G. (2008) Real-time in vivo photoacoustic and ultrasound imaging. Journal of Biomedical Optics, 13 (5). 050510. ISSN 1083-3668

[img]PDF
392Kb
Abstract:A real-time photoacoustic imaging system is designed and built. This system is based on a commercially available ultrasound imaging system. It can achieve a frame rate of 8 frames/sec. Vasculature in the hand of a human volunteer is imaged, and the resulting photoacoustic image is combined with the ultrasound image. The real-time photo acoustic imaging system with a hybrid ultrasound probe is demonstrated by imaging the branching of subcutaneous blood vessels in the hand
Item Type:Article
Copyright:©2008 Society of Photo-Optical Instrumentation Engineers