Rupture of the arterial wall causes deflection in pressure time course during ex vivo balloon angioplasty

Share/Save/Bookmark

Steenhuijsen, Jacco L.G. and Vaartjes, Simon R. and Poortermans, Cees J. and Havenith, Miek G. and Boom, Herman B.K. (1997) Rupture of the arterial wall causes deflection in pressure time course during ex vivo balloon angioplasty. Catheterization and Cardiovascular Diagnosis, 42 (1). pp. 92-101. ISSN 0098-6569

[img]
Preview
PDF
261Kb
Abstract:A relation between restenosis and arterial lesions resulting from balloon angioplasty has been suggested in literature. Nevertheless, it is unclear to what extent angioplasty-induced arterial wall lesions contribute to the occurrence of restenosis. One problem is that arterial ruptures cannot be detected during balloon inflation. This study describes a method to detect ruptures in the arterial wall, based on deflections observable in the development of the balloonpressure. We performed ex vivo angioplasty with constant strain rate on 28 human femoral artery segments, showing deflections in 21 cases. In 20 cases wall rupture was confirmed histologically. From seven cases not showing deflections, four showed intact wall at microscopy. These figures result in a selectivity of the proposed method of 87 ± 7% and a predictive value of the positive test of 95 ± 5%. We conclude that this method can enhance detection of arterial rupture during ex vivo angioplasty and may become important clinically.
Item Type:Article
Copyright:© 1997 Wiley InterScience
Research Group:
Link to this item:http://purl.utwente.nl/publications/71341
Official URL:http://dx.doi.org/10.1002/(SICI)1097-0304(199709)42:1<92::AID-CCD26>3.0.CO;2-N
Export this item as:BibTeX
EndNote
HTML Citation
Reference Manager

 

Repository Staff Only: item control page