The tonometric sensor, a new device for the measurement of intraocular pressure
Voorthuyzen, J.A. and Besten den, C. and Bergveld, P. (1989) The tonometric sensor, a new device for the measurement of intraocular pressure. Sensors and Actuators, 17 (3-4). pp. 547-552. ISSN 0250-6874
| PDF 363Kb |
| Abstract: | In this paper we present a new sensor for the measurement of intraocular pressure. We have applied the indentation principle, in which the eye is indented by exerting a force on it, while the size of the indented area is monitored. To measure the force we have used a commercial force sensor. The sensor to measure the indentation of the eye has been developed in our laboratories. We have performed measurements using an experimental set-up; these show that an accuracy of 90% will be possible if the sensor is designed in the proper way. |