Surface plasmon resonance immunosensors: sensitivity considerations
Kooyman, R.P.H. and Kolkman, H. and Gent van, J. and Greve, J. (1988) Surface plasmon resonance immunosensors: sensitivity considerations. Analytica Chimica Acta, 213 . pp. 35-45. ISSN 0003-2670
| PDF 672Kb |
| Abstract: | Some aspects of improving surface plasmon resonance response in immunosensing applications are considered. Both from calculations and experiments, it was found that maximum sensitivity is obtained for a silver layer about 55 nm thick in direct contact with the species to be quantified. Application of an intermediate layer with high permittivity can be useful in suppressing background responses. Experimentally, a protein surface-coverage fraction of ca. 0.1 could be measured, corresponding to ca. 10−10 mol1−1 antibody. |
| Item Type: | Article |
| Copyright: | © 1988 Elsevier |
| Faculty: | Science and Technology (TNW) |
| Research Group: | |
| Link to this item: | http://purl.utwente.nl/publications/70325 |
| Official URL: | http://dx.doi.org/10.1016/S0003-2670(00)81337-9 |
| Export this item as: | BibTeX EndNote HTML Citation Reference Manager |
Repository Staff Only: item control page

Show download statistics for this publication
Show download statistics for this publication