Design considerations for an isfet multiplexer and amplifier
Bergveld, Piet (1984) Design considerations for an isfet multiplexer and amplifier. Sensors and Actuators, 5 (1). pp. 13-20. ISSN 0250-6874
| PDF 615Kb |
| Abstract: | Design considerations for the multiplexing of pH-sensitive ISFETs are studied and discussed. Experimental results with an ordinary multiplex circuit show transients of the order of several seconds if the output voltage of the applied amplifier is required within an accuracy of 0.5 mV (corresponding to 0.01 pH unit). These transients appear to result from thermal instability of the ISFETs during switching operations, and can be cancelled by guaranteeing a constant chip dissipation. This necessitates the design of a special multiplexer circuit for multiplexing separate ISFETs. The output voltage of the applied amplifier then appears to be stable within 1 millisecond with the required accuracy. |
| Item Type: | Article |
| Copyright: | © 1984 Elsevier Science |
| Faculty: | Electrical Engineering, Mathematics and Computer Science (EEMCS) |
| Link to this item: | http://purl.utwente.nl/publications/69331 |
| Official URL: | http://dx.doi.org/10.1016/0250-6874(84)87002-X |
| Export this item as: | BibTeX EndNote HTML Citation Reference Manager |
Repository Staff Only: item control page
Show download statistics for this publication
Show download statistics for this publication