Improvement of the performance of a mu -metal magnetically shielded room by means of active compensation (biomagnetic applications)
Brake ter, H.J.M. and Wieringa, H.J. and Rogalla, H. (1991) Improvement of the performance of a mu -metal magnetically shielded room by means of active compensation (biomagnetic applications). Measurement Science and Technology, 2 (7). p. 596. ISSN 0957-0233
| PDF 437Kb |
| Abstract: | An active compensation technique is presented for improving the performance of a mu -metal magnetically shielded room. Active compensation is established by measuring the magnetic field inside the room by a SQUID magnetometer. The output of this sensor is amplified and connected to a coil surrounding the room. The magnetic field generated in this way compensates the measured field inside the room. Active compensation was tested for magnetic fields in the vertical direction in a shielded room with one mu -metal shield. At low frequencies a shielding improvement of typically 40 dB was obtained. Measurements performed on a room with two mu -metal shields indicated that the attainable improvement is not limited by the amount of mu -metal applied in the room. The active compensation set-up is described in detail and experiments performed on the two magnetically shielded rooms are presented and discussed. |
| Item Type: | Article |
| Copyright: | © 1991 Institute of Physics |
| Faculty: | Science and Technology (TNW) |
| Research Group: | |
| Link to this item: | http://purl.utwente.nl/publications/60690 |
| Official URL: | http://dx.doi.org/10.1088/0957-0233/2/7/004 |
| Export this item as: | BibTeX EndNote HTML Citation Reference Manager |
Repository Staff Only: item control page
Metis ID: 129144

Show download statistics for this publication
Show download statistics for this publication