A monitoring device for pressurised-air-driven diaphragm-based artificial heart assist devices
Hoeben, F.P. and Mul de, F.F.M. and Stokkink, H.S.D. and Koelink, M.H. and Greve, J. (1992) A monitoring device for pressurised-air-driven diaphragm-based artificial heart assist devices. Clinical Physics and Physiological Measurement, 13 (2). pp. 161-166. ISSN 0143-0815
| PDF 275Kb |
| Abstract: | A non-invasive device has been developed to monitor the diaphragm position and the blood flow in artificial heart assist devices equipped with a pressurised-air-driven diaphragm. Light scattering from the diaphragm is used as a mechanism for measuring. Information about the position of several points of the diaphragm can be obtained. The completely empty or filled situation can be detected and used for control purposes. Flow data can be extracted and bending characteristics of the diaphragm during operation can be studied. |
| Item Type: | Article |
| Copyright: | © Institute of Physics and IOP Publishing Limited 1992 |
| Research Group: | |
| Link to this item: | http://purl.utwente.nl/publications/59006 |
| Official URL: | http://dx.doi.org/10.1088/0143-0815/13/2/008 |