The Twente Photoacoustic Mammoscope: system overview and performance

Share/Save/Bookmark

Manohar, Srirang and Kharine, Alexei and Hespen, Johan C.G. van and Steenbergen, Wiendelt and Leeuwen, Ton G. van (2005) The Twente Photoacoustic Mammoscope: system overview and performance. Physics in Medicine and Biology, 50 (11). pp. 2543-2557. ISSN 0031-9155

[img] PDF
Restricted to UT campus only
: Request a copy
552kB
Abstract:We present PAM, the Photoacoustic Mammoscope developed at the University
of Twente, intended for initial retrospective clinical studies on subjects with
breast tumours. A parallel plate geometry has been adopted and the breast will
be gently compressed between a glass plate and a flat ultrasound detector matrix. Pulsed light (5 ns) from an Nd:YAG laser will impinge the breast through the glass plate in regions of interest; an appropriate number of the 590 elements of the detector matrix will be activated in succession to record photoacoustic signals. Three-dimensional image reconstruction employs a delay-and-sum beamforming algorithm. We discuss various instrumental aspects and the
proposed imaging protocol. Performance studies of the ultrasound detector are presented in terms of sensitivity, frequency response and resolution. Details of the patient–instrument interface are provided. Finally some imaging results on well-characterized breast tissue phantoms with embedded tumour simulating
inserts are shown.
Item Type:Article
Copyright:© Institute of Physics
Faculty:
Science and Technology (TNW)
Research Group:
Link to this item:http://purl.utwente.nl/publications/58699
Official URL:http://dx.doi.org/10.1088/0031-9155/50/11/007