Photoacoustic determination of blood vessel diameter

Share/Save/Bookmark

Kolkman, Roy G.M. and Klaessens, John H.G.M. and Hondebrink, Erwin and Hopman, Jeroen C.W. and Mul de, Frits F.M. and Steenbergen, Wiendelt and Thijssen, Johan M. and Leeuwen van, Ton G. (2004) Photoacoustic determination of blood vessel diameter. Physics in Medicine and Biology, 49 (20). pp. 4745-4756. ISSN 0031-9155

[img]PDF
Restricted to UT campus only
: Request a copy
202Kb
Abstract:A double-ring sensor was applied in photoacoustic tomographic imaging of artificial blood vessels as well as blood vessels in a rabbit ear. The peak-topeak time (τ pp) of the laser (1064 nm) induced pressure transient was used to estimate the axial vessel diameter. Comparison with the actual vessel diameter showed that the diameter could be approximated by 2cτ pp, with c the speed of sound in blood. Using this relation, the lateral diameter could also precisely be determined. In vivo imaging and monitoring of changes in vessel diameters was feasible. Finally, acoustic time traces were recorded while flushing a vessel in the rabbit ear with saline, which proved that the main contribution to the laser-induced pressure transient is caused by blood inside the vessel and that the vessel wall gives only a minor contribution.
Item Type:Article
Copyright:© 2004 Institute of Physics
Faculty:
Science and Technology (TNW)
Research Group:
Link to this item:http://purl.utwente.nl/publications/47861
Official URL:http://dx.doi.org/10.1088/0031-9155/49/20/006
Export this item as:BibTeX
EndNote
HTML Citation
Reference Manager

 

Repository Staff Only: item control page

Metis ID: 218823