Mols, R.F.X.A.M. and Ernst, G.J. (1994) Performance of a MOPA laser system for photocathode research. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 341 (1-3). pp. 481-483. ISSN 0168-9002

[img]
Preview
PDF
242Kb
Abstract:A Nd:YLF laser system with frequency doubled output is described. Several aspects concerning energy and stability of the pulses will be discussed. The system consists of a CW mode locked Nd:YLF oscillator and two double pass amplifiers having a total small signal gain of about a million which means care must be taken to prevent self-oscillations. Due to saturation effects the energy per micropulse is limited to 20 µJ, resulting in 8 µJ at the second harmonic frequency (526.5 nm). For the frequency conversion KTP type II is used. The short term timing stability was measured using a spectrum analyser and found to be less than 200 fs with a width of 450 Hz. The amplitude fluctuations over the 15 ¿s macropulse are determined by the saturation behaviour of the amplifiers and the shot to shot stability of the amplifiers. Third and fourth harmonic generation is under study.