Development of an integrated NSOM probe
Ruiter, A.G.T. and Moers, M.H.P. and Jalocha, A. and Hulst van, N.F. (1995) Development of an integrated NSOM probe. Ultramicroscopy, 61 (1-4). pp. 139-143. ISSN 0304-3991
| PDF 795Kb |
| Abstract: | A new type of NSOM probe has been developed, with a design based on the probes used in Atomic Force Microscopy. The probe consists of a cantilever with at its end a conical tip. This tip has been metal-coated to provide an aperture. With the cantilevered probe, the problem of breaking of the tip due to high normal forces is solved. In operation, the tip is scanned in contact with the sample while regulating the force between the tip and the sample with a beam deflection technique, which allows to simultaneously make an optical and a topographical image of the sample. The probes are made using micromechanical techniques, which allows batch fabrication of the probes. Testing of the probes is done in a transmission NSOM set-up in which the sample is scanned while the tip and the optical path are kept fixed. Using an opaque sample with submicron holes, the new probes have been tested, resulting an optical image with a simultaneously measured topographical image. |
| Item Type: | Article |
| Copyright: | © 1995 Elsevier Science |
| Research Group: | |
| Link to this item: | http://purl.utwente.nl/publications/24571 |
| Official URL: | http://dx.doi.org/10.1016/0304-3991(95)00119-0 |
| Export this item as: | BibTeX EndNote HTML Citation Reference Manager |