Multichannel heart scanner based on high-Tc SQUIDs

Share/Save/Bookmark

Brake, H.J.M. ter and Janssen, N. and Flokstra, J. and Veldhuis, D. and Rogalla, H. (1997) Multichannel heart scanner based on high-Tc SQUIDs. IEEE Transactions on Applied Superconductivity, 7 (2). pp. 2545-2548. ISSN 1051-8223

open access
[img]
Preview
PDF
588kB
Abstract:A 7-channel magnetometer for magnetocardiography based on high-T c SQUIDs has been realized. This magnetometer is used for test experiments in the development of a multichannel high-Tc SQUID based heart-scanner for clinical applications. The intrinsic noise level of the channels in the 7-channel system is typically 120 fT/¿(Hz) down to 1 Hz. Magnetocardiograms were recorded inside a magnetically shielded room. Introductory experiments were performed on the suppression of noise by combining magnetometers to form planar gradiometers. The noise suppression that can be established appeared to be limited by the imbalance of the gradiometric configuration, which is roughly 2%. This relatively poor balance of the system is caused by inaccuracies in the transfer functions of the individual SQUID magnetometers, and by deviations from the planar geometry
Item Type:Article
Copyright:©1997 IEEE
Faculty:
Science and Technology (TNW)
Research Group:
Link to this item:http://purl.utwente.nl/publications/24469
Official URL:http://dx.doi.org/10.1109/77.621758
Export this item as:BibTeX
EndNote
HTML Citation
Reference Manager

 

Repository Staff Only: item control page

Metis ID: 129269