Analysis of current density and related parameters in spinal cord stimulation

Share/Save/Bookmark

Wesselink, Wilbert A. and Holsheimer, Jan and Boom, Herman B.K. (1998) Analysis of current density and related parameters in spinal cord stimulation. IEEE Transactions on Rehabilitation Engineering, 6 (2). pp. 200-207. ISSN 1063-6528

open access
[img]
Preview
PDF
151kB
Abstract:A volume conductor model of the spinal cord and surrounding anatomical structures is used to calculate current (and current density) charge per pulse, and maximum charge density per pulse at the contact surface of the electrode in the dorsal epidural space, in the dorsal columns of the spinal cord and in the dorsal roots. The effects of various contact configurations (mono-, bi-, and tripole), contact area and spacing, pulsewidth and distance between contacts and spinal cord on these electrical parameters were investigated under conditions similar to those in clinical spinal cord stimulation. At the threshold stimulus of a large dorsal column fiber, current density and charge density per pulse at the contact surface were found to be highest (1.9·105 ¿A/cm2 and 39.1 ¿C/cm2 ·p, respectively) when the contact surface was only 0.7 mm 2. When stimulating with a pulse of 500 ¿s, highest charge per pulse (0.92 ¿C/p), and the largest charge density per pulse in the dorsal columns (1.59 ¿C/cm2·p) occurred. It is concluded that of all stimulation parameters that can be selected freely, only pulsewidth affects the charge and charge density per pulse in the nervous tissue, whereas both pulsewidth and contact area strongly affect these parameters in the nonnervous tissue neighboring the electrode contacts
Item Type:Article
Copyright:© 1998 IEEE
Research Group:
Link to this item:http://purl.utwente.nl/publications/15415
Official URL:http://dx.doi.org/10.1109/86.681186
Export this item as:BibTeX
EndNote
HTML Citation
Reference Manager

 

Repository Staff Only: item control page

Metis ID: 112290