Perception threshold and electrode position for spinal cord stimulation

Share/Save/Bookmark

He, Jiping and Barolat, Giancarlo and Holsheimer, Jan and Struijk, Johannes J. (1994) Perception threshold and electrode position for spinal cord stimulation. Pain, 59 (1). pp. 55-63. ISSN 03043959

[img]
Preview
PDF
1017Kb
Abstract:The perception threshold for epidural spinal cord stimulation in chronic pain management was analyzed on 3923 testing data obtained from 136 implanted patients. The initial areas of paresthesiae due to stimulation were recorded and reported as the stimulation map according to the location of electrodes. Measurement of dorsal thickness of the cerebrospinal fluid (CSF) layer was obtained from 26 subjects using magnetic resonance imaging (MRI). The results indicate that the perception threshold is a function of the spinal level of the implanted electrodes, of the mediolateral position in the spinal canal and the contact separation of electrode. Differences in perception threshold at various vertebral levels are mainly due to varying depths of the dorsal CSF layer. The medially placed electrodes caudal to the mid-cervical levels have a higher perception threshold than more laterally placed ones. The electrodes at high and mid-cervical levels, however, have a smaller perception threshold if placed medially. The information obtained from this investigation has important implications for the design of a new-generation stimulation system and clinical application to maximize the longevity of the power source.
Item Type:Article
Copyright:© 1994 Elsevier Science
Faculty:
Electrical Engineering, Mathematics and Computer Science (EEMCS)
Research Group:
Link to this item:http://purl.utwente.nl/publications/15360
Official URL:http://dx.doi.org/10.1016/0304-3959(94)90047-7
Export this item as:BibTeX
EndNote
HTML Citation
Reference Manager

 

Repository Staff Only: item control page

Metis ID: 112235