Evanescent wave sensing: new features for detection in small volumes
Pandraud, G. and Koster, T.M. and Gui, C. and Dijkstra, M. and Berg van den, A. and Lambeck, P.V. (2000) Evanescent wave sensing: new features for detection in small volumes. Sensors and Actuators A: Physical, 85 (1-3). pp. 158-162. ISSN 0924-4247
| PDF Restricted to UT campus only: Request a copy 196Kb |
| Abstract: | The association by aligned direct bonding of an evanescent wave optical sensor and small-volume flow cells for visible absorption detection is demonstrated. We show that for sample volumes ranging from 760 to 190 nl and a sensor interaction length of 1 mm, very small cross-sectional dimensions of fluid channels are no more a serious obstacle for optical detection. The lowest limit for the channel depth is the penetration depth of the |
| Item Type: | Article |
| Copyright: | © 2000 Elsevier |
| Faculty: | Electrical Engineering, Mathematics and Computer Science (EEMCS) |
| Research Group: | |
| Link to this item: | http://purl.utwente.nl/publications/15207 |
| Official URL: | http://dx.doi.org/10.1016/S0924-4247(00)00367-8 |
| Export this item as: | BibTeX EndNote HTML Citation Reference Manager |
Repository Staff Only: item control page
Metis ID: 112082

Show download statistics for this publication
Show download statistics for this publication