Automated injection of slurry samples in flow-injection analysis

Share/Save/Bookmark

Hulsman, M. and Bos, M. and Linden van der, W.E. (1996) Automated injection of slurry samples in flow-injection analysis. Analytica Chimica Acta, 324 (1). pp. 13-19. ISSN 0003-2670

[img]
Preview
PDF
636Kb
Abstract:Two types of injectors are described for introducing solid samples as slurries in flow analysis systems. A time-based and a volume-based injector based on multitube solenoid pinch valves were built, both can be characterized as hydrodynamic injectors. Reproducibility of the injections of dispersed solids ( 150 ¿m) was tested with several concentrations of slurry samples up to 30 mg/ml; the injected volume was 1 ml. For both injectors dye and slurry samples could be injected with good precision (relative standard deviation for the peak area less than 2%). Peak detection was performed turbidimetrically. Data analysis and operation of the injectors were automated. The usual peristaltic pumps in flow analysis are normally not capable of handling slurries of the type investigated, therefore a valveless piston pump was used instead.
Item Type:Article
Copyright:© 1996 Elsevier Science
Research Group:
Link to this item:http://purl.utwente.nl/publications/12536
Official URL:http://dx.doi.org/10.1016/0003-2670(96)00005-0
Export this item as:BibTeX
EndNote
HTML Citation
Reference Manager

 

Repository Staff Only: item control page

Metis ID: 106717